IMMULITE 2000 XPi Immunoassay System

The highly reliable IMMULITE® 2000 XPi system is easy to use and has one of the largest automated immunoassay menus available. Medium- to high-volume laboratories can incorporate specialty and allergy testing into routine workloads to improve productivity and efficiency.

  • Customized automated startup for daily routine maintenance and quality control scheduling
  • Reagent onboard stability of 90 days
  • Tube-top sample cups for micro-sample handling
  • Home screens with at-a-glance status on a 19 inch touch screen monitor
  • Options available for direct water feed and bulk solid waste chute
  • Siemens Automation solutions available through VersaCell® Systems

Features & Benefits

The IMMULITE 2000 XPi system is a continuous random access immunoassay analyzer with a maximum throughput of 200 tests per hour. This design increases workflow and enhances productivity for medium- to high-volume immunoassay laboratories by incorporating specialty and allergy testing into one platform. 

The IMMULITE 2000 XPi system has the reliability expected from the IMMULITE family of products.

IMMULITE® 2000 XPi Immunoassay System

Enhancing the Laboratory workflow

  • 90 day onboard stability of reagents minimizes waste
  • Maximum throughput of 200 tests per hour 
  • No pause sample loading to enhance turnaround time
  • Re-run, dilution and reflex capabilities to minimize technologist time
  • AutoStart functionality, including automated daily maintenance and quality control scheduling, reduces hands-on operations

 

IMMULITE 2000XPi sample cup

Improving efficiency through the Intuitive and Easy to use Software

  • 19 inch touch screen monitor for easy navigation
  • Color coded consumable levels always available at a glance
  • User configurable dilution and reflex testing to meet laboratory needs

 

Providing flexible solutions

  • Cost effective automation available through connectivity through VersaCell® solutions
  • Scalable connectivity to Siemens track based automation solutions

Assays

Allergy
AlaTOP Allergy Screen*
Allergen-Specific IgE 3gAllergy™
Allergen-Specific IgG*
Allergen-Specific IgG4*
ECP*
IgE, Total  

 

Anemia
EPO*
Ferritin
Folate
RBC Folate
Vitamin B12

 

Autoimmune
Anti-CCP

 

Bone Metabolism
Osteocalcin*
Pyrilinks-D

Cardiac
CKMB*
D-Dimer*
High-Sensitivity CRP
Myoglobin
NT-proBNP*
Troponin I

Diabetes
C-Peptide
Insulin
Microalbumin

Growth
Growth Hormone (hGH)
IGF-I
IGFBP-3

Hepatitis
HBsAb
HBsAb Quantitative
HBc IgM
HBc Total
HBsAg
HBsAg Confirmatory

 

Inflammation
IL-2R*
IL-6*
LBP*

 

Metabolic
ACTH
Cortisol
Homocysteine

Oncology

AFP
CA 125 (OM-MA)
CA 15-3 (BR-MA)
CA 19-9* (GI-MA)
Calcitonin
CEA
Free PSA
PAP
PSA

Other
ß2-Microglobulin
Gastrin
Nicotine Metabolites*

Reproductive Endocrinology
Androstenedione
DHEAS
Estradiol
Free Beta HCG*
FSH
hCG
LH
PAPP-A*
Progesterone
Prolactin
SHBG
Testosterone
Unconjugated Estriol

Special ID
EBV-EBNA IgG*
EBV-VCA IgG*
EBV-VCA IgM*
H. pylori IgG
Syphilis

Therapeutic Drug Monitoring
(TDM)
Carbamazepine
Digitoxin*
Digoxin
Phenobarbital
Phenytoin
Theophylline
Valproic Acid

Thyroid
Anti-TG
Anti-TPO
Free T3
Free T4
Intact PTH
T Uptake
Thyroid Binding Globulin (TBG)
Thyroglobulin (Tg)
Total T3
Total T4
TSH
TSH, 3rd Generation
TSI (Thyroid Stimulating Immunoglobin)

 

ToRCH
CMV IgG
CMV IgM
Herpes I & II IgG
Rubella IgG
Rubella IgM
Toxoplasma IgG
Toxoplasma IgM


Veterinary
Canine TLI
Canine Total T4
Canine TSH
Canine Free T4

Technical Specifications

Overview

System Description

Random access immunoassay system

Throughput

Up to 200 tests per hour in batch or random access mode

Time To First Result

35 minutes

Assays On Board

24

Samples and Sample Handling

Sample Tubes

12-16 mm diameter
75-100 mm heights
10 x 50 microtubes with tube-top racks

Sample Carriers

15 position sample racks; 90 samples fully loaded; Continuous loading

Walk Away Time

Up to 5 hours

Sample Types

Serum, plasma, urine – assay dependent

Sample Integrity Control

Pressure-transducer technology with clot detection, clot flagging, and short sample detection

Automatic Sample
Processing

Reflex, re-run, and dilutions

Barcodes

Formats Supported

Code 39, Code 128, Codabar, UPC,and I-2 of 5 (Interleaved)

Reagents and Reagent Handling

Reagent Positions

24

Reagent Cooling

2 - 8° C

Reagent Preparation

None required

Reagent On Board Sability

90 days

Reagent Barcodes

Allows system to identify lot number, and determine when calibrator adjustment needed, and the tests remaining in the reagent wedge

User Interface

Operating System

Microsoft® Windows® 7

Monitor

19 inch touch sensitive screen

Maintenance

Daily

Automated daily maintenance cycle completed according to time programmed by the laboratory

Weekly

20 minutes

Monthly

45 minutes

Environmental

Operating Temperature

18 – 32° C

Electrical

 

200 - 240 v, 6 AMP
50/60 Hz

Dimensions

Height

65 in / 165 cm

Width

93 in / 236cm

Depth

45 in / 114 cm

Weight