ADVIA 120 Hematology SystemMinimizing review rates and repeats utilizing proven flow cytometry

The ADVIA® 120 System offers a unique portfolio of diagnostic assays and gold-standard flow cytometry testing. To satisfy the demands of laboratories, Siemens Healthineers delivers a hematology analyzer with comprehensive testing parameters, flexible sample handling, a user-defined interface, and a compact bench-top design.

  • Automates hematology workflow without the need for large track-based systems or reflexive testing
  • Maximizes efficiency through user-defined screen configurations and reports
  • Simplifies maintenance with Unifluidics Technology through reduced fluidics, elimination of pinch valves, and automated daily cleaning
  • Streamlines productivity with a 150 capacity rack-based autosampler, allowing 75 minutes of true walk-away time
  • Reduces the need for manual slide production and review by utilizing WBC and RBC/PLT technology and high specificity
  • Maximizes the effectiveness of costly platelet transfusions with accurate results the first time—even at very low platelet levels

Features & Benefits

The ADVIA® 120 Hematology System can automate your hematology lab without the need for large track-based systems or reflexive testing. This is how Siemens delivers practical solutions to your hematology lab. 

Streamline workflow and maximize productivity with a hematology analyzer that eliminates the majority of commonly performed manual steps. The ADVIA 120 System delivers the gold-standard in testing methodology for optimum results while offering the simplicity and flexibility you need for easy integration into your lab. 

ADVIA technology utilizes peroxidase staining—the gold-standard in differential testing. This primary methodology of the ADVIA Hematology Systems also provides a secondary total white cell count that acts as an internal QC check to monitor sample integrity.

  • Unique cytochemical differentiation of myeloid and lymphoid cells, and six-parameter differential supplies additional clinical information
  • Direct measurements of intracellular hemoglobin provide sensitive and specific RBC morphology for the accurate and early indication of disease states
  • Accurate platelet counts in the clinically significant range (excluding microcytic red cells and cell fragments) minimize the number of manual counts and unnecessary costly platelet transfusions
  • Further isolate and identify disease states without additional labor through CSF and CHr testing capabilities

Maximize your workflow efficiency with gold-standard technology. With integrated and automated solutions from Siemens, you can scale from stand-alone instrumentation to complete automation as your productivity needs change.

Assays

Parameters                     

 

CBC Results

WBC, RBC, HGB†, HCT, MCV, MCH, MCHC, CHCM, RDW, HDW, CH, CHDW, PLT

Differential Results
(absolute and %)

NEUT, LYMPH, MONO, EOS, BASO, LUC (Large Unstained Cells)

Platelet Results

PLT, MPV, PDW, PCT

Reticulocyte Results
(absolute and %)

RETIC, MCVr, CHCMr, RDWr**, HDWr**, CHr, CHDWr**

CSF assay results

Optional

Morphology results
(User-definable)

 

WBC

Left Shift, Atypical Lymph, Blasts, Immature Granulocytes, Myeloperoxidase Deficiency

RBC and PLT

NRBC, ANISO, MICRO, MACRO, HC VAR, HYPO, HYPER, RBC Fragments, RBC Ghosts, Platelet Clumps, Large Platelets

Technical Specifications

WBC

0.02 to 400 x 103/µL

RBC

0.0 to 7.0 x 106/µL

PLT

5.0 to 3500 x 103/µL

HB

0.0 to 22.5 g/dL

Retic

0.2 to 24.5%

Parameter

Mean

SD

CV

WBC

7.5

0.20

2.7

RBC

5.0

0.06

1.2

HB

15.0

0.14

0.93

MCV

90.0

0.7

0.78

PLT

300.0

8.8

2.93

Retic%

2.0

0.25

12.5

CBC

120 samples/hour

CBC/DIFF

120 samples/hour

CBC/DIFF/Retic

74 samples/hour

CBC/Retic

74 samples/hour

Retic

74 samples/hour

Automatic Closed Tube

175µL

Manual Closed Tube

175µL

Manual Open Tube

175µL

Analytical Module with Autosampler:

Weight

193 kg / 425 lbs

Height

86 cm / 33.8 in

Width

141 cm / 55.5 in

Depth

68 cm / 26.8 in

Analytical Module without Autosampler:

Weight

163 kg / 360 lbs

Height

86 cm / 33.8 in

Width

81 cm / 31.9 in

Depth

68 cm / 26.8 in

Intel Pentium E6500 Dual Core 2.93 GHz
350W Desktop/Tower
3Gb/s 7200 RPM 16MB Cache Hard Drive
1 GB Memory Module
CD-RW
17” Color Monitor
Full-sized Keyboard (major languages)

150 Sample Capacity:

15 Racks with 10 tube capacity

Compatible tube types:

Vacutainer®, Hemogard™, Monovette® (Sarstedt 2.6 mL tubes require rack modification), Greiner MicroSample Tubes, BD MAP tubes

Automatic discrimination of label codes:

Codabar
Interleave 2 of 5 with and without check digit
Code 39
Code128 EAN and JAN (8 and 13)

Onboard analytical and wash reagents (1850 CBC/Diff tests)
Automated Daily Maintenance – fully programmable