| Random access immunoassay system |
| |
| 18 minutes; results every 15 seconds thereafter, assay dependent |
| |
| ADVIA Centaur® ReadyPack™ assays |
| Direct Chemiluminescence using Acridinium Ester technology |
| |
| 1 mL, 2 mL, 3 mL, 5 mL, 7 mL, 10 mL, micro container tubes |
| Universal 5-position rack holds multiple tube types; 180 samples fully loaded; no-pause loading and unloading |
| Dedicated STAT port accepts samples any time |
| Serum, plasma, urine - assay dependent |
| 10 uL to 200 uL assay dependent |
| Pressure transducer technology with clot management, detection, and flagging. |
| Ensure zero sample carryover |
| Varies by assay, up to 1:1000 |
| |
| Code 128, Code 39, Codabar, Interleaved 2 of 5 |
| |
| 30-position reagent tray cooled 4°C to 8°C |
Reagent Integrity Control | Barcode reagent identification, automatic inventory tracking and flagging, calibration validity tracked and flagged, reagent on board stability tracked and flagged, reagent expired/ reagent low flagging |
| ReadyPacks automatically rocked onboard |
| 25-position ancillary reagent tray cooled 4°C to 8°C |
| |
| 1,000 ADVIA Centaur cuvette capacity |
| ADVIA Centaur type, 840 on board, automatic tracking, and flagging of inventory |
| |
| Loading/unloading of samples and consumables at anytime without he need to interrupt system operations, includes: sample racks, STATS, reagent packs, ancillary reagent packs, reaction cuvettes, sample pipette tips, calibrators, controls, wash fluids, water, and waste. |
| |
| 19-inch diagonal high-resolution LCD touch screen with adjustable height |
Operating System Results Management | Dual Sun Microsystems Sparc® II processors; Microsoft® Windows® XP |
| RS232C bidirectional and host query; Ethernet connection available; 25,000 result storage; archive to removable media; automatic reflex testing |
| Repeats and confirms reactive testing; results available as completed |
| |
| Advanced QC package includes Levy-Jennings plots and Westgard rules; 65,000 control results can be stored |
| |
| |
| |
| 100 Base-T Ethernet port for remote access and Siemens Remote Service proactive service modem |
| |
| 208-240 V at 50/60 Hz; meets regulatory requirements for UL, CAN/CSA, IEC1010. JIS, CE, CISPR11 Class A |
| 1 KW in ready state; 1.5 KW in operating state |
| |
| 51.5 (h) x 72.4 (w) x 41.0 (d) in-130.9 (h) x 183.9 (w) x 104.2 (d) cm |
| |
Floor Load-bearing Requirement | |
Operating Environment Temperature | 18-30OC; humidity 20-80% noncondensing |
| |
| 58.8 db in ready state; 61.3 in operating state |