MAGNETOM Vida – the first 3T MRI scanner with BioMatrix Technology – embraces human nature and the unique set of challenges that each patient brings to the MRI exam. By embracing these challenges, we overcome patient variability – ushering in a paradigm shift in MRI. The outcome: fewer rescans, predictable scheduling and consistent, high-quality personalized exams for high-end clinical routine as well as clinical research.
MAGNETOM VidaEmbrace human nature at 3T
Features & Benefits
Embrace full 3T performance with unparalleled magnet and gradient power
3T MRI is synonymous with high-performance clinical MRI. With 3T MRI, the expectation is to push and redefine the limits of what is possible in diagnostic imaging.
An all-new 3T magnet
MAGNETOM Vida comes with an all-new 3T magnet, which provides excellent homogeneity throughout the entire measurement volume and a large Field-of-View of 55 x 55 x 50 cm3. The result: fast examinations with robust and reliable fat saturation throughout the entire imaging volume, which is especially important in abdominal or off-center applications.
A new level of gradient power
With a gradient strength of up to 60/200 simultaneously for up to 25% higher SNR in diffusion weighted imaging, MAGNETOM Vida offers MR research power – but now in a patient-friendly 70 cm setting.
Cost-efficient energy management
An intelligent power management disables power-consuming components during scan breaks – EcoPower shuts off the cold head in standby, resulting in up to 30% energy savings compared to the industry average (COCIR average of MRI vendors3,4).
Embrace true 3T productivity with Turbo Suite and GO Technologies
MAGNETOM Vida embraces true 3T productivity and makes fast scan times and push-button examinations become clinical reality. At the core of its impressive acquisition speed lies our exclusive Turbo Suite: Disruptive acceleration technologies (e.g. Simultaneaous Multi-Slice and Compressed Sensing) shortening the entire examinations for the whole spectrum of clinical MRI, while keeping the quality of exam results.
GO Technologies are a family of solutions that accelerates the entire workflow of the examination: from patient positioning to result distribution.
GO Technologies consist of four main components, each contributing to a specific aspect of workflow optimization to enable higher throughput and robustness in clinical routine.
Embrace new 3T clinical capabilities with Inline Compressed Sensing
With Turbo Suite Elite, MAGNETOM Vida provides access to patients previously excluded from MRI because of their medical condition or the exam’s prohibitively complicated nature. Compressed Sensing Cardiac Cine and Compressed Sensing GRASP-VIBE enable free-breathing exams which help improve the patient experience, while extending your range of services. Significantly simplified procedures with automated workflow strengthen clinical fields with growth potential such as whole-body MRI.
With MR Fingerprinting (MRF)5, MAGNETOM Vida enables diagnostics at the frontier of a paradigm shift. MRF uses quantitative information from scans for decisions based on digital tissue data, not visual impressions, to improve tissue differentiation and enable less invasive diagnostics.
Clinical Use
See the consistent high-quality perzonalized exams of MAGNETOM Vida and our BioMatrix technology.
Full whole-body coverage in only 5 stations
- BioMatrix Head/Neck 20
- 2 x Body 18
- BioMatrix Spine 72
- Peripheral Angio 36
Courtesy of University Hospital Tübingen, Germany
SMS RESOLVE
SMS RESOLVE
PAT 2 SMS 2, b1000 | 1.0 x 1.0 x 4.0 mm3 | TA 2:32 min
- BioMatrix Head/Neck 64
Customer contribution from Austria
3D CS SPACE DIR
Left:
3D SPACE DIR | PAT 2 | 1.4 mm iso | TA 6:07 min
Right:
3D CS SPACE DIR | CS 7 | 1.0 x 1.0 x 1.0 mm3 (interpolated) | TA 3:07 min
- BioMatrix Head/Neck 20
DTI fibertracking 64 directions, post-processed with Neuro 3D
- BioMatrix Head/Neck 64
MR Fingerprinting – It’s about writing the next chapter in MR
- BioMatrix Head/Neck 20
C-spine imaging with and without CoilShim
- BioMatrix Head/Neck 20 (18°tilt)
BioMatrix SliceAdjust
- BioMatrix Spine 32
Hip imaging with the new UltraFlex coils
- UltraFlex Large 18
Courtesy of Radiologie München Harlaching, Munich, Germany
Knee imaging – with the new Tx/Rx Knee 18
- Tx/Rx Knee 18
Courtesy of University Hospital Tübingen, Germany
Foot/Ankle imaging with the new UltraFlex coils
- UltraFlex Large 18
Courtesy of Radiologie München Harlaching, Munich, Germany
Shoulder imaging
Left:
PD TSE FS | PAT 2 | 0.4 x 0.4 x 3.0 mm3 | TA 4:06 min
Right:
SMS PD TSE FS | PAT 2 SMS 2 | 0.4 x 0.4 x 3.0 mm3 | TA 3:08 min
- Shoulder Shape 16
Abdominal imaging with the new Contour 24 Coil
- Contour 24 Coil
3D T1 CAIPIRINHA VIBE Dixon
50 cm z-FOV
Full abdominal coverage from liver dome down to the pelvis allow for comprehensive oncological scans in one station with excellent quality.
- Body 18
- BioMatrix Spine 72
High-resolution MRCP
3D T2 SPACE | PAT 3 | 0.5 x 0.5 x 1.0 mm3 | TA 5:05 min resp. triggered
Middle:
3D T2 CS SPACE | CS 20 | 0.5 x 0.5 x 1.0 mm3 | TA 1:26 min resp. triggered
Right:
3D T2 CS SPACE | CS 23 | 0.4 x 0.4 x 1.6 mm3 | TA 0:15 min breath-hold
- Body 18
- BioMatrix Spine 32
Compressed Sensing GRASP-VIBE
- Body 18
- BioMatrix Spine 72
Courtesy of University Hospital Tübingen, Germany
BioMatrix Respiratory Sensor
T2 TSE fBLADE FS
- BioMatrix Spine 72
- Body 18
MR Angiography of the carotids
- BioMatrix Head/Neck 20
- BioMatrix Spine 32
3aaaa0156
Compressed Sensing Cardiac Cine
Left:
- BioMatrix Spine 32
- Body 18
- BioMatrix Spine 72
1 Courtesy of Benson Radiology, Australia
2 Courtesy of University Hospital Erlangen, Germany
Brachial plexus – 3D T2 SPACE
- BioMatrix Head/Neck 64
- UltraFlex Large 18
Full whole-body coverage in only 5 stations
- BioMatrix Head/Neck 20
- 2 x Body 18
- BioMatrix Spine 72
- Peripheral Angio 36
Courtesy of University Hospital Tübingen, Germany
SMS RESOLVE
SMS RESOLVE
PAT 2 SMS 2, b1000 | 1.0 x 1.0 x 4.0 mm3 | TA 2:32 min
- BioMatrix Head/Neck 64
Customer contribution from Austria
3D CS SPACE DIR
Left:
3D SPACE DIR | PAT 2 | 1.4 mm iso | TA 6:07 min
Right:
3D CS SPACE DIR | CS 7 | 1.0 x 1.0 x 1.0 mm3 (interpolated) | TA 3:07 min
- BioMatrix Head/Neck 20
DTI fibertracking 64 directions, post-processed with Neuro 3D
- BioMatrix Head/Neck 64
MR Fingerprinting – It’s about writing the next chapter in MR
- BioMatrix Head/Neck 20
C-spine imaging with and without CoilShim
- BioMatrix Head/Neck 20 (18°tilt)
BioMatrix SliceAdjust
- BioMatrix Spine 32
Hip imaging with the new UltraFlex coils
- UltraFlex Large 18
Courtesy of Radiologie München Harlaching, Munich, Germany
Knee imaging – with the new Tx/Rx Knee 18
- Tx/Rx Knee 18
Courtesy of University Hospital Tübingen, Germany
Foot/Ankle imaging with the new UltraFlex coils
- UltraFlex Large 18
Courtesy of Radiologie München Harlaching, Munich, Germany
Shoulder imaging
Left:
PD TSE FS | PAT 2 | 0.4 x 0.4 x 3.0 mm3 | TA 4:06 min
Right:
SMS PD TSE FS | PAT 2 SMS 2 | 0.4 x 0.4 x 3.0 mm3 | TA 3:08 min
- Shoulder Shape 16
Abdominal imaging with the new Contour 24 Coil
- Contour 24 Coil
3D T1 CAIPIRINHA VIBE Dixon
50 cm z-FOV
Full abdominal coverage from liver dome down to the pelvis allow for comprehensive oncological scans in one station with excellent quality.
- Body 18
- BioMatrix Spine 72
High-resolution MRCP
3D T2 SPACE | PAT 3 | 0.5 x 0.5 x 1.0 mm3 | TA 5:05 min resp. triggered
Middle:
3D T2 CS SPACE | CS 20 | 0.5 x 0.5 x 1.0 mm3 | TA 1:26 min resp. triggered
Right:
3D T2 CS SPACE | CS 23 | 0.4 x 0.4 x 1.6 mm3 | TA 0:15 min breath-hold
- Body 18
- BioMatrix Spine 32
Compressed Sensing GRASP-VIBE
- Body 18
- BioMatrix Spine 72
Courtesy of University Hospital Tübingen, Germany
BioMatrix Respiratory Sensor
T2 TSE fBLADE FS
- BioMatrix Spine 72
- Body 18
MR Angiography of the carotids
- BioMatrix Head/Neck 20
- BioMatrix Spine 32
3aaaa0156
Compressed Sensing Cardiac Cine
Left:
- BioMatrix Spine 32
- Body 18
- BioMatrix Spine 72
1 Courtesy of Benson Radiology, Australia
2 Courtesy of University Hospital Erlangen, Germany
Brachial plexus – 3D T2 SPACE
- BioMatrix Head/Neck 64
- UltraFlex Large 18
Full whole-body coverage in only 5 stations
- BioMatrix Head/Neck 20
- 2 x Body 18
- BioMatrix Spine 72
- Peripheral Angio 36
Courtesy of University Hospital Tübingen, Germany
See the full application range in the MAGNETOM Vida Image Gallery.
Hear from the experts
"In our large enterprise inconsistency just kills us. So I´m really impressed with the slice by slice shim for body diffusion, the coil shim for fat suppression, for things like brachial plexus and spine and soft tissue neck. [They] give me the predictability that I need to plan a concise day and be patient centric in our imaging."
Lawrence Tanenbaum, M.D.
Radnet, New York, USA
"To provide our patients with individual therapies, we need every piece of information available. In imaging, this means we need robust, standardized, and reproducible image data, always of the same quality regardless of the patient or user. MAGNETOM Vida with BioMatrix Technology gives us this data quality and comprehensive image information and is helping us on our way to quantitative radiology."
Professor Konstantin Nikolaou, M.D.
University Hospital Tübingen, Germany
"MAGNETOM Vida with Compressed Sensing Cardiac Cine and PSIR HeartFreeze enables us to scan patients without breathing commands. We have compared the free-breathing technology to a conventional breath-hold approach and beyond time advantage, patients are more compliant and no diagnostic information is lost."
Associate Professor Wolfgang Wüst, M.D.
University Hospital Erlangen, Germany
"MAGNETOM Vida brings the performance of a true research system – large FoV and very strong gradients and combines it with a 70 cm bore which is, of course, more comfortable for patients."
Professor Mike Notohamiprodjo, M.D.
University Hospital Tübingen, Germany
"Compressed Sensing GRASP-VIBE establishes completely new capabilities in abdominal imaging. Being able to perform DCE studies under free breathing not only increases the robustness of liver MRI but also generates data which can be used for pharmacokinetic modelling. GRASP-VIBE is very useful for many applications [...] due to its inherent robustness towards involuntary motion."
Sergios Gatidis, M.D.
University Hospital Tübingen, Germany
Clinical acceleration: from the console
Read how Benson Radiology (North Adelaide, Australia) with their new MAGNETOM Vida speed up imaging without compromising on image quality and spatial resolution.
Find more MAGNETOM Vida specific clinical talks and case studies on MAGNETOM World, the Siemens MR user community forum.
Technical Details
Magnet System |
|
Field strength | 3 Tesla |
Bore size | 70 cm / 2.30 ft Open Bore design |
Helium consumption | Zero Helium boil-off technology |
Shimming | Passive and active |
|
|
Gradient strength | XQ gradients 45/200 simultaneously [2.03 MW] |
|
|
Power consumption | Eco-Power technology |
|
|
Siting and Installation |
|
System length | 186 cm / 6.10 ft cover to cover |
System weight (in operation) | 7.35 tons / 15,432.36 lb |
Minimum room size7 | 31 m2 / 333.68 sq ft |