| Random-access clinical chemistry and immunoassay system with LOCI® chemiluminescence technology |
| Performs up to 440 photometric chemistry tests per hour and 187 IMT tests per hour on serum, plasma, urine, and cerebrospinal fluid; can perform up to 167 heterogeneous immunoassay tests per hour
|
| IMT (NA, K, CL): 1.5 min; Electrolytes (IMT + ECO2): 2.9 min; BMP-7 Electrolytes, GLU, BUN, CREA) 4.0 min; CMP-6 (ALB, DBIL, TBIL, AST, ALT, ALP): 9.3 min; TNI: 11 min; HCG: 15.5 min |
| |
Disease-state Assay Groups | Anemia, Autoimmune/Rheumatoid, Bone Metabolism, Cardiovascular, Diabetes, Drugs of Abuse/Toxicology, Hepatic Diseases, Immunosuppressive Drugs/TDM, Inflammation, Nephropathies, Nutritional Assessment, Pancreatic Disease, Oncology, Reproductive Endocrinology, Thyroid |
| |
| |
| 5 mL, 7 mL, 10 mL tubes; 1.5 mL sample cups; 1.0 mL small sample containers and pediatric tubes |
| 60 sample positions in six 10-tube segments; positive sample identification |
| Not dedicated; STAT samples are processed with priority |
| Qualitative check for hemolysis, lipemia, and icterus; clot detection, flagging, and management; short-sample detection, flagging, and management
|
| Automatic repeat testing from the original sample
|
| |
| Automatic dilution: 1:1.5 up to 1:200 |
| Automatic reflex testing based on results of first test
|
| Liquid-level sensing, clot detection, short-sample detection
|
Sample Carryover Prevention | |
| Up to 200 tubes per hour as part of the VersaCell® X3 Solution – Dimension® Suite; faster on automation |
| |
| |
| Code 39; Code 128; Codabar (USS); Interleaved 2 of 5 w and w/o check digit, 12 digits maximum |
| |
| |
| Onboard capacity of 12,000 formed cuvettes
|
| Air; incubation temperature 37°C
|
| |
| The filter wheel holds optical filters for wavelengths of 293, 340, 383, 405, 452, 510, 540, 577, 600, 680, and 700 nm
|
| Standard tungsten halogen lamp, operation at 6.5A (6.8v)
|
| 1, 3, 4, 5, 10, 15, 21, and 32 minutes
|
| Serum blank, cell blank, reagent blank, measurement point change, autodilution
|
| LOCI, heterogeneous immunoassay, PETINIA and ACMIA, photometry, potentiometry (ISE), turbidimetric, and Emit® |
Assay Result Calculations | Endpoint, rate, multipoint |
| |
| |
| 2 trays, 44 positions, refrigerated between 2–8°C (36–47°F); one on the Reagent Management System (RMS)
|
| 88 Flex® Reagent Cartridges plus 3 electrolytes via the QuikLYTE® IMT
|
| 3 probes with liquid-level sensing
|
| Flex Reagent Cartridges, bar coded, 15 to 360 tests/Flex
|
Average Total Reaction Volume
| |
| Bar-code reagent identification; automatic inventory tracking and flagging; calibration and control validity tracking and flagging; reagent onboard tracking of tests remaining, lot number, onboard stability, and expiration date
|
| Depending on assay, up to 42 days |
| 25,200 tests average; 33,300 tests maximum |
| |
| |
| 110 assay channels; includes 10 channels for user-defined applications
|
| |
| |
| Indirect simultaneous measurement of Na+, K+, Cl-
|
| 40 μL for all three tests
|
| Automatic priming cycle, no user calibration, automatic urine dilution 1:10
|
| 1000 samples or 5 days, whichever comes first
|
| |
| |
Validated Calibration Interval | Up to 90 days, tracked by software, with 500 most recent calibration logs stored electronically if a system restore is required (starting with software version 10.1)
|
| Assay-specific time interval or with new reagent lot
|
| User-defined time interval |
| Graphical display of calibration curves |
| Graphical display of QC plot (histogram or Levey-Jennings) with Westgard Rules; RealTime QC; QCC PowerPak™ efficiency package
|
| |
User Interface/Data Management
| |
| 17-inch diagonal touchscreen with adjustable height |
| |
| |
| 100,000 patient tests (10 MB), 100,000 QC results (10 MB), 9000 calibrations (5 yrs, 18 MB) |
| |
| |
| System requests work order or batch of work orders from host
|
| Ethernet port for remote access via Siemens Remote Service (starting with software version 10.1); modem for remote diagnostic access |
| |
| |
| CD, DVD, and USB (starting with software version 10.0) |
| |
| |
| 115 VAC at 60 Hz (nominal); 11 amps max; 1.3 kW consumption in operating state 230 VAC at 50 Hz (nominal); 5.5 amps max; 1.3 kW consumption in operating state |
| 115 VAC at 60 Hz (nominal); 4 amps max; 450 W consumption in operating state 230 VAC at 50 Hz (nominal); 3 amps max; 450 W consumption in operating state |
| - Instrument feed pressurized water source
- <3.8 bar (<55 psi)
- Instrument feed water system must maintain
- stable DO2 content between 5 and 8 ppm²
- Temperature: <35°C (<95°F)
- Resistivity: >10 megohms cm
- Bacterial content: <10 colony forming units/mL
- System feed water line must not exceed 3 m (12 feet)
|
| - Instrument may be supplied with a water purifier that provides instrument feed water
- If an alternative water system is used, water must adhere to Siemens water specifications
|
| |
| |
| Normal operation: 252 cm W x 122 cm H x 132 cm D (99 in. W x 48 in. H x 52 in. D) With monitor fully extended, lids fully open, and external UPS: 255 cm W x 191 cm H x 155 cm D (100 in. W x 75 in. H x 61 in. D)
|
| |
| Complies with international environmental, health, and safety standards |
| <75 dBA at 1 m while operating
|
| |
Operating Temperature Range
| |
| |